作者
J Kühn, F Charrière, T Colomb, Etienne Cuche, F Montfort, Yves Emery, P Marquet, Christian Depeursinge
发表日期
2008/5/23
期刊
Measurement Science and Technology
卷号
19
期号
7
页码范围
074007
出版商
IOP Publishing
简介
We present state-of-the-art dual-wavelength digital holographic microscopy (DHM) measurement on a calibrated 8.9 nm high chromium thin step sample and demonstrate sub-nanometer axial accuracy. By using a modified DHM reference calibrated hologram (RCH) reconstruction method, a temporal averaging procedure and a specific dual-wavelength DHM arrangement, it is shown that specimen topography can be measured with an accuracy, defined as the axial standard deviation, reduced to at least 0.9 nm. Indeed for the first time to the best of our knowledge, it is reported that averaging each of the two wavefronts recorded with real-time dual-wavelength DHM can provide up to 30% spatial noise reduction for the given configuration. Moreover, the presented experimental configuration achieves a temporal stability below 0.8 nm, thus paving the way to Angström range for dual-wavelength DHM.
引用总数
20092010201120122013201420152016201720182019202020212022202320246181651117109971369867
学术搜索中的文章
J Kühn, F Charrière, T Colomb, E Cuche, F Montfort… - Measurement Science and Technology, 2008