简介
Objectives
To compare technical and clinical screening performance parameters between computed radiography (CR) and direct digital radiography (DR) systems.
Methods
The number of women screened with CR was 73,008 and with DR 116,945. Technical and patient dose survey data of 25 CR and 37 DR systems were available. Technical performance was expressed by threshold thickness values at the mean glandular dose (MGD) level of routine practice. Clinical indicators included recall rate (RR), cancer detection rate (CDR), percentage of ductal carcinoma in situ (DCIS), percentage of cancers with T-scores smaller than 1 cm and positive predictive value (PPV).
Results
Contrast threshold values for the 0.1-mm gold disk were 1.44 μm (SD 0.13 μm) for CR and 1.20 μm (SD 0.13 μm for DR). MGD was 2.16 mGy …