作者
Osvaldo Rampado, Alessandro Depaoli, Filippo Marchisio, Marco Gatti, Damien Racine, Valeria Ruggeri, Irene Ruggirello, Fatemeh Darvizeh, Paolo Fonio, Roberto Ropolo
发表日期
2021/1
期刊
La radiologia medica
卷号
126
页码范围
55-62
出版商
Springer Milan
简介
Purpose
The purpose of this study was to verify the maintenance of low-contrast detectability at different CT dose reduction levels, in patients of different sizes, as a consequence of the application of iterative reconstruction at different strengths combined with tube current modulation.
Methods
Anthropomorphic abdominal phantoms of two sizes (small and large) were imaged at a fixed noise with iterative algorithm ASIR-V percentages in the range between 0 and 70% and corresponding dose reductions in the range of 0–83%. A total of 1400 images with and without liver low-contrast simulated lesions were evaluated by five radiologists, using the receiver operating characteristics (ROC) paradigm and evaluating the area under the ROC curve (AUC). The human observer results were then compared with AUC obtained with a channelized Hotelling observer …