作者
Amir T Moinzadeh, Lydia Farack, Fabian Wilde, Hagay Shemesh, Paul Zaslansky
发表日期
2016/5/1
期刊
Journal of endodontics
卷号
42
期号
5
页码范围
776-781
出版商
Elsevier
简介
Introduction
This study evaluated the integrity of calcium silicate sealer–based fillings made with hygro-expandable cones (HEC) that are commercially known as CPoint or Smartpoint.
Methods
Fourteen human canines were prepared according to a standardized, conventional endodontic treatment protocol and filled with the HEC/calcium silicate sealer. Three-dimensional imaging was performed with laboratory micro–computed tomography (μCT) at its highest resolution and was compared with synchrotron phase contrast-enhanced μCT (PCE-CT) scans of the treatment extending 1–7 mm from the apex. Conventional destructive optical microscopy validated observations by comparison with virtual slices in the tomographic data.
Results
Conventional laboratory μCT at 10-μm resolution did not reveal the existing voids and defects within the root canal fillings. PCE-CT revealed elongated interfacial delamination …
引用总数
20162017201820192020202120222023202421172421