Background
Contact force (CF) information may improve the safety and efficacy of ablation for paroxysmal atrial fibrillation (PAF).
Objective
The purpose of this study was to assess the impact of CF data on ablation for PAF.
Methods
Patients undergoing first-time PAF ablation were randomized at 7 UK centers to ablation with (CF-on) or without (CF-off) CF data available to the operator, using the same ablation catheter and mapping system. An ablation CF of 5–40g was targeted. Pulmonary vein (PV) reconnection was assessed with adenosine at 60 minutes. Follow-up for arrhythmia recurrence was for 1 year with 7-day Holter recordings at 6 and 12 months.
Results
One hundred seventeen patients were studied (59 CF-on, 58 CF-off). In the CF-on group, a reduction in acute PV reconnection rates (22% vs 32%, P = .03) but no significant difference in 1-year success rates off antiarrhythmic drugs (49% vs 52%, P = .9) was …