BACKGROUND & AIMS
Three-dimensional (3D) ultrasound imaging of the total stomach volume has not yet been achieved. The aim of this study was to investigate whether a magnetic position sensor system for acquisition of 3D ultrasonograms could be used to determine gastric emptying rates and intragastric distribution.
METHODS
A system for position and orientation measurement was interfaced to an ultrasound scanner. In vitro accuracy was evaluated by scanning a porcine stomach. Fourteen volunteers, with a median age of 35 years, were scanned fasting and postcibally by two-dimensional (2D) and 3D ultrasound after ingesting a 500-mL soup meal.
RESULTS
This 3D system yielded a strong correlation (r = 0.997) between true and estimated volumes in vitro. The limits of agreement were -9.1:70.1 mL in the volume range 1200-1900 mL. The intersubject variability of the total gastric volumes ranged from …