作者
J Greffier, J Frandon, A Larbi, JP Beregi, F Pereira
发表日期
2020/1
期刊
European radiology
卷号
30
页码范围
487-500
出版商
Springer Berlin Heidelberg
简介
Purpose
To assess the dose performance in terms of image quality of filtered back projection (FBP) and two generations of iterative reconstruction (IR) algorithms developed by the most common CT vendors.
Materials and methods
We used four CT systems equipped with a hybrid/statistical IR (H/SIR) and a full/partial/advanced model-based IR (MBIR) algorithms. Acquisitions were performed on an ACR phantom at five dose levels. Raw data were reconstructed using a standard soft tissue kernel for FBP and one iterative level of the two IR algorithm generations. The noise power spectrum (NPS) and the task-based transfer function (TTF) were computed. A detectability index (d′) was computed to model the detection task of a large mass in the liver (large feature; 120 HU and 25-mm diameter) and a small calcification (small feature; 500 HU and 1.5-mm …