作者
Tianye Niu, Mingshan Sun, Josh Star‐Lack, Hewei Gao, Qiyong Fan, Lei Zhu
发表日期
2010/10
期刊
Medical physics
卷号
37
期号
10
页码范围
5395-5406
出版商
American Association of Physicists in Medicine
简介
Purpose
Applications of cone‐beam CT (CBCT) to image‐guided radiation therapy (IGRT) are hampered by shading artifacts in the reconstructed images. These artifacts are mainly due to scatter contamination in the projections but also can result from uncorrected beam hardening effects as well as nonlinearities in responses of the amorphous silicon flat panel detectors. While currently, CBCT is mainly used to provide patient geometry information for treatment setup, more demanding applications requiring high‐quality CBCT images are under investigation. To tackle these challenges, many CBCT correction algorithms have been proposed; yet, a standard approach still remains unclear. In this work, we propose a shading correction method for CBCT that addresses artifacts from low‐frequency projection errors. The method is consistent with the current workflow of radiation therapy.
Methods
With much smaller …