Ruben Pauwels, Jilke Beinsberger, Bruno Collaert, Chrysoula Theodorakou, Jessica Rogers, Anne Walker, Lesley Cockmartin, Hilde Bosmans, Reinhilde Jacobs, Ria Bogaerts, Keith Horner, SEDENTEXCT Project Consortium
发表日期
2012/2/1
期刊
European journal of radiology
卷号
81
期号
2
页码范围
267-271
出版商
Elsevier
简介
OBJECTIVE
To estimate the absorbed organ dose and effective dose for a wide range of cone beam computed tomography scanners, using different exposure protocols and geometries.
MATERIALS AND METHODS
Two Alderson Radiation Therapy anthropomorphic phantoms were loaded with LiF detectors (TLD-100 and TLD-100H) which were evenly distributed throughout the head and neck, covering all radiosensitive organs. Measurements were performed on 14 CBCT devices: 3D Accuitomo 170, Galileos Comfort, i-CAT Next Generation, Iluma Elite, Kodak 9000 3D, Kodak 9500, NewTom VG, NewTom VGi, Pax-Uni3D, Picasso Trio, ProMax 3D, Scanora 3D, SkyView, Veraviewepocs 3D. Effective dose was calculated using the ICRP 103 (2007) tissue weighting factors.
RESULTS
Effective dose ranged between 19 and 368μSv. The largest contributions to the effective dose were from the remainder tissues (37 …
引用总数
201120122013201420152016201720182019202020212022202320241859729086748178646565463713
学术搜索中的文章
R Pauwels, J Beinsberger, B Collaert, C Theodorakou… - European journal of radiology, 2012