作者
Kyung Hwan Chang, Dong Wook Kim, Jung Hoan Choi, Han-Back Shin, Chae-Seon Hong, Dong Min Jung, Joo Ho Kim, Jihun Kim, Mincheol Han, Ho Lee, Kwangwoo Park, Jin Sung Kim, Jaeho Cho, Yong Bae Kim, Joo Ah Lee
发表日期
2020/2
期刊
Journal of the Korean Physical Society
卷号
76
页码范围
257-263
出版商
The Korean Physical Society
简介
In this study, the delivery quality assurance (DQA) results of commercially available dosimetric systems (ionization chamber and EBT film, MapCHECK, ArcCHECK, and dosimetry check (DC) software) for helical tomotherapy (HT) were compared, and the feasibility of performing pretreatment using MapCHECK, ArcCHECK, and DC for HT, instead of ionization chambers and EBT films, was assessed. Sixty-five HT-treated patients were considered. Point dose differences, dose profiles, and gamma passing rates were used to evaluate the agreement between the calculated and the measured data, and the outcomes of the four DQA devices were compared in various clinical cases. The calculated and the measured point doses were within ±5% of each other. In terms of the root-mean-square error (RMSE), the point dose differences were within 2.9 for the four tested devices in all of the studied cases. Gamma analysis …