简介
Severe contamination of the gastric signal in electrogastrogram (EGG) analysus by respiratory, motion, cardiac artifacts, and possible myoelectrical activity from other organs, poses a major challenge to EGG interpretation and analysis. A generally applicable method for removing a variety of artifacts from EGG recordings is proposed based on the empirical mode decomposition (EMD) method. This decomposition technique is adaptive, and appears to be uniquely suitable for nonlinear, non-stationary data analysis. The results show that this method, combined with instantaneous frequency analysis, effectively separate, identify and remove contamination from a wide variety of artifactual sources in EGG recordings.