Gianfranco Mitacchione, Marco Schiavone, Alessio Gasperetti, Gianmarco Arabia, Alexander Breitenstein, Manuel Cerini, Pietro Palmisano, Elisabetta Montemerlo, Matteo Ziacchi, Simone Gulletta, Francesca Salghetti, Giulia Russo, Cinzia Monaco, Patrizio Mazzone, Daniel Hofer, Fabrizio Tundo, Giovanni Rovaris, Antonio Dello Russo, Mauro Biffi, Ennio CL Pisanò, Gian Battista Chierchia, Paolo Della Bella, Carlo De Asmundis, Ardan M Saguner, Claudio Tondo, Giovanni B Forleo, Antonio Curnis
发表日期
2023/3/1
期刊
Heart Rhythm
卷号
20
期号
3
页码范围
395-404
出版商
Elsevier
简介
Background
Limited data on the real-world safety and efficacy of leadless pacemakers (LPMs) post–transvenous lead extraction (TLE) are available.
Objective
The purpose of this study was to assess the long-term safety and effectiveness of LPMs following TLE in comparison with LPMs de novo implantation.
Methods
Consecutive patients who underwent LPM implantation in 12 European centers joining the International LEAdless PacemakEr Registry were enrolled. The primary end point was the comparison of LPM-related complication rate at implantation and during follow-up (FU) between groups. Differences in electrical performance were deemed secondary outcomes.
Results
Of the 1179 patients enrolled, 15.6% underwent a previous TLE. During a median FU of 33 (interquartile range 24–47) months, LPM-related major complications and all-cause mortality did not differ between groups (TLE group: 1.6% and 5.4 …