作者
Yi‐Fen Yen, Karl F Stupic, Michael T Janicke, Douglas N Greve, Azma Mareyam, Jason Stockmann, Jonathan R Polimeni, Andre Van Der Kouwe, Kathryn E Keenan
发表日期
2023/5
期刊
NMR in Biomedicine
卷号
36
期号
5
页码范围
e4873
简介
T1 relaxation times of the 14 T1 phantom spheres that make up the standard International Society for Magnetic Resonance in Medicine (ISMRM)/National Institute of Standards and Technology (NIST) system phantom are reported at 7 T. T1 values of six of the 14 T1 spheres at 7 T (with T1 > 270 ms) have been reported previously, but, to the best of our knowledge, not all of the T1s of the 14 T1 spheres at 7 T have been reported before. Given the increasing number of 7‐T MRI systems in clinical settings and the increasing need for T1 phantoms that cover a wide range of T1 relaxation times to evaluate rapid T1 mapping techniques at 7 T, it is of high interest to obtain accurate T1 values for all the ISMRM/NIST T1 spheres at 7 T. In this work, T1 relaxation time was measured on a 7‐T MRI scanner using an inversion‐recovery spin‐echo pulse sequence and derived by curve fitting to a signal equation that exhibits …
引用总数
学术搜索中的文章
YF Yen, KF Stupic, MT Janicke, DN Greve, A Mareyam… - NMR in Biomedicine, 2023