作者
Kevin L Schauer, Carolyn J Broccardo, Kimberly M Webb, Paul A Covey, Jessica E Prenni
发表日期
2013/7
期刊
Journal of Biomolecular Techniques: JBT
卷号
24
期号
2
页码范围
57
出版商
The Association of Biomolecular Resource Facilities
简介
The superior sensitivity of current mass spectrometers makes them prone to contamination issues, which can have deleterious effects on sample analysis. Here, bis (2, 2, 6, 6-tetramethyl-4-piperidyl) sebacate (marketed under the name Tinuvin 770) is identified as a major contaminant in applications using liquid chromatography coupled with mass spectrometry (LC-MS). Tinuvin 770 is often added to laboratory and medical plastics as a UV stabilizer. One particular lot of microcentrifuge tubes was found to have an excess of this compound that would leach into samples and drastically interfere with LC-MS data acquisition. Further analysis found that Tinuvin 770 readily leached into polar and nonpolar solvents from the contaminated tube lot. Efforts to remove Tinuvin 770 from contaminated samples were unsuccessful. A prescreening method using MALDI-TOF MS is presented to prevent system contamination and …
引用总数
201520162017201820192020202120222023202411322
学术搜索中的文章
KL Schauer, CJ Broccardo, KM Webb, PA Covey… - Journal of Biomolecular Techniques: JBT, 2013