作者
Tuncay Küme, Barıs Sağlam, Cem Ergon, Ali Rıza Sisman
发表日期
2018/1
期刊
Journal of clinical laboratory analysis
卷号
32
期号
1
页码范围
e22168
简介
Background
The aim of this study is to evaluate and compare the analytical performance characteristics of the two creatinine methods based on the Jaffe and enzymatic methods.
Methods
Two original creatinine methods, Jaffe and enzymatic, were evaluated on Architect c16000 automated analyzer via limit of detection (LOD) and limit of quantitation (LOQ), linearity, intra‐assay and inter‐assay precision, and comparability in serum and urine samples. The method comparison and bias estimation using patient samples according to CLSI guideline were performed on 230 serum and 141 urine samples by analyzing on the same auto‐analyzer.
Results
The LODs were determined as 0.1 mg/dL for both serum methods and as 0.25 and 0.07 mg/dL for the Jaffe and the enzymatic urine method respectively. The LOQs were similar with 0.05 mg/dL value for both serum methods, and enzymatic urine method had a lower …