Gerrit L ten Kate, Guillaume GJ Renaud, Zeynettin Akkus, Stijn CH van den Oord, Folkert J ten Cate, Vijay Shamdasani, Rob R Entrekin, Eric JG Sijbrands, Nico de Jong, Johan G Bosch, Arend FL Schinkel, Antonius FW van der Steen
发表日期
2012/4/1
期刊
Ultrasound in medicine & biology
卷号
38
期号
4
页码范围
593-600
出版商
Elsevier
简介
The present study describes the presence of pseudoenhancement during contrast-enhanced ultrasound (CEUS) imaging of human carotid arteries and the reproduction of this pseudoenhancement in vitro. Seventy patients underwent bilateral CEUS examination of the carotid arteries using a Philips iU22 ultrasound system equipped with a L9-3 ultrasound probe and SonoVue microbubble contrast. During CEUS of the carotid arteries, we identified enhancement in close proximity to the far wall, parallel to the main lumen. The location of this enhancement does not correlate to the anatomical location of a parallel vessel. To corroborate the hypothesis that this is a pseudoenhancement artifact, the enhancement was recreated in a tissue-mimicking material phantom, using the same ultrasound system, settings and contrast agent as the patient study. The phantom study showed that pseudoenhancement may be present …