作者
W Scott Hoge, Kawin Setsompop, Jonathan R Polimeni
发表日期
2018/10
期刊
Magnetic resonance in medicine
卷号
80
期号
4
页码范围
1364-1375
简介
Purpose
A ghost correction strategy for Simultaneous Multi‐Slice (SMS) EPI methods that provides improved ghosting artifact reduction compared to conventional methods is presented. Conventional Nyquist ghost correction methods for SMS‐EPI rely on navigator data that contain phase errors from all slices in the simultaneously acquired slice‐group. These navigator data may contain spatially nonlinear phase differences near regions of B0 inhomogeneity, which violates the linear model employed by most EPI ghost correction algorithms, resulting in poor reconstructions.
Methods
Dual‐Polarity GRAPPA (DPG) was previously shown to accurately model and correct both spatially nonlinear and 2D phase errors in conventional single‐slice EPI data. Here, an extension we call Dual‐Polarity slice‐GRAPPA (DPsG) is adapted to the slice‐GRAPPA method and applied to SMS‐EPI data for slice separation and ghost …
学术搜索中的文章
WS Hoge, K Setsompop, JR Polimeni - Magnetic resonance in medicine, 2018