作者
Gang Ye, Hua-Gen Mo, Ze-Hou Wang, Shan-Hong Yi, Xiang-Wei Wang, Yin-Fu Zhang
发表日期
2006/2/1
期刊
The Journal of urology
卷号
175
期号
2
页码范围
636-640
出版商
No longer published by Elsevier
简介
PURPOSE
The metal ring pin stapler was initially developed for microvascular surgery and there has been little experience of their use in larger vessels. We determined if the titanium ring pin coupling system could be safely and rapidly applied for arterial reconstruction in clinical renal transplantation.
MATERIALS AND METHODS
The donor renal artery was end-to-end anastomosed to the internal iliac artery with titanium ring pin staplers in 36 patients. Anastomotic, clamp and total operative time and vascular problems were compared with those in the control group of sutured anastomosis in 39 transplant recipients.
RESULTS
The completion of mechanical anastomosis required half the time of suture anastomosis. Mean clamp time was 17.8 minutes in the nonsuture group and 28.1 minutes in the control group. There was no significant difference in total operative time between the 2 groups. A small anastomotic line …