作者
Wei Zhao, Pedro L Esquinas, Xinchi Hou, Carlos F Uribe, Marjorie Gonzalez, Jean-Mathieu Beauregard, Yuni K Dewaraja, Anna Celler
发表日期
2018/12
期刊
EJNMMI physics
卷号
5
页码范围
1-16
出版商
Springer International Publishing
简介
Background
Camera calibration, which translates reconstructed count map into absolute activity map, is a prerequisite procedure for quantitative SPECT imaging. Both planar and tomographic scans using different phantom geometries have been proposed for the determination of the camera calibration factor (CF). However, there is no consensus on which approach is the best. The aim of this study is to evaluate all these calibration methods, compare their performance, and propose a practical and accurate calibration method for SPECT quantitation of therapeutic radioisotopes. Twenty-one phantom experiments (Siemens Symbia SPECT/CT) and 12 Monte Carlo simulations (GATE v6.1) using three therapy isotopes (131I, 177Lu, and 188Re) have been performed. The following phantom geometries were used: (1) planar scans of point source in air (PS), (2) tomographic scans of insert(s …
引用总数
20182019202020212022202320245671410105
学术搜索中的文章