作者
Gad Wollstein, David F Garway-Heath, Roger A Hitchings
发表日期
1998/8/1
期刊
Ophthalmology
卷号
105
期号
8
页码范围
1557-1563
出版商
Elsevier
简介
PURPOSE
This study aimed to define the confocal laser scanning ophthalmoscope (Heidelberg Retina Tomograph [HRT]) parameters that best separate patients with early glaucoma from normal subjects.
STUDY DESIGN
A cross-sectional study.
PARTICIPANTS
A total of 80 normal subjects and 51 patients with early glaucoma participated (average visual field mean deviation = −3.6 dB).
INTERVENTION
Imaging of the optic nerve head with the HRT and analysis using software version 1.11 were performed.
MAIN OUTCOME MEASURES
The relation between neuroretinal rim area and optic disc area, and cup-disc area ratio and optic disc area, was defined by linear regression of data derived from the normal subjects. The normal ranges for these two parameters were defined by the 99% prediction intervals of the linear regression between the parameter and optic disc area, for the whole disc, and for each of the …
引用总数
1999200020012002200320042005200620072008200920102011201220132014201520162017201820192020202120222023202481917113028434242444440272726302513121311911641