作者
PA Tonnu, T Ho, K Sharma, E White, C Bunce, D Garway-Heath
发表日期
2005/7/1
期刊
British journal of ophthalmology
卷号
89
期号
7
页码范围
847-850
出版商
BMJ Publishing Group Ltd
简介
Aim: To compare the inter-method agreement in intraocular pressure (IOP) measurements made with four different tonometric methods.
Methods: IOP was measured with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT) in a randomised order in one eye of each of 105 patients with ocular hypertension or glaucoma. Three measurements were made with each method, and by each of two independent GAT observers. GAT interobserver and tonometer inter-method agreement was assessed by the Bland-Altman method. The outcome measures were 95% limits of agreement for IOP measurements between GAT observers and between tonometric methods, and 95% confidence intervals for intra-session repeated measurements.
Results: The mean differences (bias) in IOP measurements were 0.4 mm Hg between GAT …