Joshua P Cysyk, Branka Lukic, Clark Joseph Brian, Ray Newswanger, Choon-Sik Jhun, Jenelle Izer, Heidi Flory, John Reibson, Bradley Doxtater, William Weiss, Gerson Rosenberg
发表日期
2021/11/1
期刊
ASAIO Journal
卷号
67
期号
11
页码范围
1240-1249
出版商
LWW
简介
We have miniaturized and optimized our implantable rotary blood pump developed to provide long-term mechanical right heart support for patients who have failing Fontan circulation. The objective of this study was to evaluate the miniaturized Fontan circulation assist device (mini-FCAD) during 30-day sheep studies (n= 5). A complete right heart bypass was performed and all return flow was supported by the pump. Postoperatively, unfractionated heparin was given to maintain thromboelastography R times of 2× normal. The first two studies were terminated on day 0 and day 4 due to complications. In the final three studies, the animals remained healthy and were electively terminated at 30±2 days. Pump flow was between 5 and 7 lpm, left atrial pressure remained normal, and inlet pressures were between 3 and 18 mm Hg with no incidents of suction. There was no evidence of hemolysis, end organ or pulmonary …