作者
Alexander Klein, Bernd Wegener, Andreas Büttner, Carolin Melcher, Volkmar Jansson, Christoph von Schulze Pellengahr
发表日期
2022/2/1
期刊
Clinical Biomechanics
卷号
92
页码范围
105589
出版商
Elsevier
简介
Background
Sufficient primary stability is mandatory for successful bony prosthetic incorporation. Therefore, defined micromotion values of 150 μm should not be exceeded as higher values might compromise the ingrowth of bone trabeculae to the implant surface. The aim of this study was to evaluate the primary stability of different cementless disc prosthesis in a cadaver model.
Methods
Four different implants with different anchoring and bearing concept were tested with a target level of L4/5. 26 specimens were randomly allocated to 1 of the 4 different implants with 6 speciments in each group. Two groups were formed depending on the anchoring (spikes vs. fin) and bearing concept (non−/semi- vs. constrained). Each implant was tested regarding primary stability in a hydraulic simulator allowing simultaneous polyaxial segment movements and axial loading. The measurements were recorded on the lower plate of …