作者
Roy M John, Sunil Kapur, Kenneth A Ellenbogen, Jayanthi N Koneru
发表日期
2017/2/1
期刊
Heart Rhythm
卷号
14
期号
2
页码范围
184-189
出版商
Elsevier
简介
Background
Collateral damage has been reported with use of the cryoballoon for pulmonary vein isolation.
Objective
The purpose of this study was to determine the incidence and characteristics associated with atrioesophageal fistula (AEF) after cryoballoon use.
Methods
Cases of AEF reported with use of the cryoballoon since 2011 were collected from the Manufacturer and User Facility Device Experience (MAUDE) database, publications, and the manufacturer’s database. Lowest balloon temperatures were compared with matched control patients undergoing cryoballoon ablation without AEF formation. Location of AEF was compared with AEF associated with radiofrequency ablation.
Results
A total of 11 cases of AEF were identified from a worldwide experience that exceed 120,000 cases. Mean age was 60 (range 31–78 years), and 80% of patients were male. Although mean lowest balloon temperatures were no …