Martin S Judenhofer, Ciprian Catana, Brian K Swann, Stefan B Siegel, Wulf-Ingo Jung, Robert E Nutt, Simon R Cherry, Claus D Claussen, Bernd J Pichler
发表日期
2007/9
期刊
Radiology
卷号
244
期号
3
页码范围
807-814
出版商
Radiological Society of North America
简介
Purpose: To prospectively use compact avalanche photodiodes instead of photomultiplier tubes to integrate a positron emission tomographic (PET) detector and a 7-T magnetic resonance (MR) imager.
Materials and Methods: All animal experiments were performed in accordance with the University of Tübingen guidelines and the German law for the protection of animals. A compact lutetium oxyorthosilicate–avalanche photodiode PET detector was built and optimized to operate within a 7-T MR imager. The detector performance was investigated both outside and inside the magnet, and MR image quality was evaluated with and without the PET detector. Two PET detectors were set up opposite each other and operated in coincidence to acquire PET images in the step-and-shoot mode in a mouse head specimen after injection of fluorine 18 fluorodeoxyglucose.
Results: The performance of the PET detector when …