作者
Ahsan Javed, Rajiv Ramasawmy, Kendall O’Brien, Christine Mancini, Pan Su, Waqas Majeed, Thomas Benkert, Himanshu Bhat, Anthony F Suffredini, Ashkan Malayeri, Adrienne E Campbell‐Washburn
发表日期
2022/4
期刊
Magnetic resonance in medicine
卷号
87
期号
4
页码范围
1784-1798
简介
Purpose
To develop an isotropic high‐resolution stack‐of‐spirals UTE sequence for pulmonary imaging at 0.55 Tesla by leveraging a combination of robust respiratory‐binning, trajectory correction, and concomitant‐field corrections.
Methods
A stack‐of‐spirals golden‐angle UTE sequence was used to continuously acquire data for 15.5 minutes. The data was binned to a stable respiratory phase based on superoinferior readout self‐navigator signals. Corrections for trajectory errors and concomitant field artifacts, along with image reconstruction with conjugate gradient SENSE, were performed inline within the Gadgetron framework. Finally, data were retrospectively reconstructed to simulate scan times of 5, 8.5, and 12 minutes. Image quality was assessed using signal‐to‐noise, image sharpness, and qualitative reader scores. The technique was evaluated in healthy volunteers, patients with coronavirus disease …
引用总数
学术搜索中的文章
A Javed, R Ramasawmy, K O'Brien, C Mancini, P Su… - Magnetic resonance in medicine, 2022