作者
James D Sanders, Dmitry Grinfeld, Konstantin Aizikov, Alexander Makarov, Dustin D Holden, Jennifer S Brodbelt
发表日期
2018/4/2
期刊
Analytical chemistry
卷号
90
期号
9
页码范围
5896-5902
出版商
American Chemical Society
简介
We demonstrate a method for determining the collision cross-sections (CCSs) of protein ions based on the decay rate of the time-domain transient signal from an Orbitrap mass analyzer. Multiply charged ions of ubiquitin, cytochrome c, and myoglobin were generated by electrospray ionization of both denaturing solutions and ones with high salt content to preserve native-like structures. A linear relationship between the pressure in the Orbitrap analyzer and the transient decay rate was established and used to demonstrate that the signal decay is primarily due to ion-neutral collisions for protein ions across the entire working pressure range of the instrument. The CCSs measured in this study were compared with previously published CCS values measured by ion mobility mass spectrometry (IMS), and results from the two methods were found to differ by less than 7% for all charge states known to adopt single gas …
引用总数
2018201920202021202220232024210471155
学术搜索中的文章