作者
Christopher M Rose, Matthew JP Rush, Nicholas M Riley, Anna E Merrill, Nicholas W Kwiecien, Dustin D Holden, Christopher Mullen, Michael S Westphall, Joshua J Coon
发表日期
2015/6/26
期刊
Journal of the American Society for Mass Spectrometry
卷号
26
期号
11
页码范围
1848-1857
出版商
Springer US
简介
Electron transfer dissociation (ETD) has been broadly adopted and is now available on a variety of commercial mass spectrometers. Unlike collisional activation techniques, optimal performance of ETD requires considerable user knowledge and input. ETD reaction duration is one key parameter that can greatly influence spectral quality and overall experiment outcome. We describe a calibration routine that determines the correct number of reagent anions necessary to reach a defined ETD reaction rate. Implementation of this automated calibration routine on two hybrid Orbitrap platforms illustrate considerable advantages, namely, increased product ion yield with concomitant reduction in scan rates netting up to 75% more unique peptide identifications in a shotgun experiment.
引用总数
201520162017201820192020202120222023202423846116571
学术搜索中的文章
CM Rose, MJP Rush, NM Riley, AE Merrill… - Journal of the American Society for Mass Spectrometry, 2015