作者
Junlan Lu, Ziyi Wang, Elianna Bier, Suphachart Leewiwatwong, David Mummy, Bastiaan Driehuys
发表日期
2022/8
期刊
Magnetic resonance in medicine
卷号
88
期号
2
页码范围
802-816
简介
Purpose
To correct for RF inhomogeneity for in vivo 129Xe ventilation MRI using flip‐angle mapping enabled by randomized 3D radial acquisitions. To extend this RF‐depolarization mapping approach to create a flip‐angle map template applicable to arbitrary acquisition strategies, and to compare these approaches to conventional bias field correction.
Methods
RF‐depolarization mapping was evaluated first in digital simulations and then in 51 subjects who had undergone radial 129Xe ventilation MRI in the supine position at 3T (views = 3600; samples/view = 128; TR/TE = 4.5/0.45 ms; flip angle = 1.5; FOV = 40 cm). The images were corrected using newly developed RF‐depolarization and templated‐based methods and the resulting quantitative ventilation metrics (mean, coefficient of variation, and gradient) were compared to those resulting from N4ITK correction.
Results
RF‐depolarization and template …