作者
Carlos Ubeda, Eliseo Vano, Patricia Miranda, Fernando Leyton, Luis Carlos Martinez, Carlos Oyarzun
发表日期
2011/11/1
期刊
Radiation protection dosimetry
卷号
147
期号
3
页码范围
429-438
出版商
Oxford University Press
简介
Radiation dose and image quality for paediatric protocols in all five X-ray fluoroscopy systems used for interventional cardiology procedures existing in Chile have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object (TO) and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4–16 cm of PMMA). Images from fluoroscopy (low (FL), medium and high) and cine (CI) modes have been archived in DICOM format. Signal-to-noise ratio (SNR), figure of merit (FOM) and high-contrast spatial resolution (HCSR) have been computed from the images. The ratio between the maximum and the minimum value of ESAK per frame for a given fluoroscopy mode between the five systems ranges from 2 to 5 and from 14 to 38 for CI mode. SNR, FOM and HCSR showed a great variability for the different acquisition modes (AMs) and …
学术搜索中的文章
C Ubeda, E Vano, P Miranda, F Leyton, LC Martinez… - Radiation protection dosimetry, 2011