作者
Grégoire P Chatain, Alton Oldham, Juan Uribe, Bradley Duhon, Michael J Gardner, Jens-Peter Witt, Scott Yerby, Brian P Kelly
发表日期
2023/10/28
期刊
Journal of Orthopaedic Surgery and Research
卷号
18
期号
1
页码范围
807
出版商
BioMed Central
简介
Background
Iliosacral screw placement is ubiquitous and now part of the surgeon’s pelvic trauma armamentarium. More recent evidence supports sacroiliac arthrodesis for treating sacroiliac joint (SIJ) dysfunction in select patients. Regardless of the surgical indication, there are currently no studies examining lag screw compression biomechanics across the SIJ. The objective of this biomechanical investigation was to quantify iliosacral implant compressive loads and to examine the insertion torque and compressive load profile over time.
Methods
Eight human cadaveric pelvic specimens underwent SIJ fixation at S1 and S2 using 11.5 and 10.0 mm iFuse-TORQ Lag implants, respectively, and standard 7.3 mm trauma lag screws. Load decay analysis was performed, and insertion and removal torques were measured.
Results
For both implants at S1 and S2 levels, the load relaxed 50% in approximately 67 min …
学术搜索中的文章
GP Chatain, A Oldham, J Uribe, B Duhon, MJ Gardner… - Journal of Orthopaedic Surgery and Research, 2023