作者
Tim Dumbleton, Arjan W Buis, Angus McFadyen, Brendan F McHugh, Geoff McKay, Kevin D Murray, Sandra Sexton
发表日期
2009/3/1
期刊
Journal of Rehabilitation Research & Development
卷号
46
期号
3
简介
In this study, we investigated and compared the dynamic interface pressure distribution of hands-off and handson transtibial prosthetic systems by means of pressure mapping. Of the 48 established unilateral amputees recruited, half (n= 24) had been wearing pressure-cast prostheses (IceCast Compact) and the other half (n= 24) had been wearing hand-cast sockets of the patellar tendon bearing design. We measured the dynamic pressure profile of more than 90% of the area within each prosthetic socket by means of four Tekscan F-Scan socket transducer arrays. We compared the interface pressure between socket concepts. We found that the distribution of dynamic pressure at the limb-socket interface was similar for the two intervention (socket prescription) groups. However, a significant difference was found in the magnitude of the interface pressure between the two socket concepts; the interface pressures recorded in the hands-off sockets were higher than those seen in the hands-on concept. Despite the differences in interface pressure, the level of satisfaction with the sockets was similar between subject groups. The sockets instrumented for this study had been in daily use for at least 6 months, with no residual-limb health problems.
引用总数
2011201220132014201520162017201820192020202120222023202449101511121211101611378
学术搜索中的文章
T Dumbleton, AW Buis, A McFadyen, BF McHugh… - Journal of Rehabilitation Research & Development, 2009