作者
Alessandro Schneebeli, Deborah Falla, Ron Clijsen, Marco Barbero
发表日期
2020/2/1
期刊
BMJ open sport & exercise medicine
卷号
6
期号
1
页码范围
e000726
出版商
BMJ Specialist Journals
简介
Objective
This study evaluates the intra-rater and inter-rater reliability of the MyotonPRO and its construct validity for the assessment of Achilles tendon stiffness.
Design
Reliability and construct validity study.
Methods
Forty healthy participants were assessed using the MyotonPRO by two raters on two different occasions. Tendon was evaluated in three different positions (relaxed, 0° plantarflexion and standing) and during different isometric contractions (range 0–3 kg). Reliability was calculated using intraclass correlation coefficient (ICC and 95% CI) standard error of measurement and minimal detectable change. Construct validity was evaluated between the different positions and the different contraction intensities using Friedman test.
Results
Intra-rater reliability was very high ICC2,k 0.87–0.98. The reliability of the 0.5 kg contraction was moderate with an ICC2,k of 0.59. Inter-rater reliability ranged from high to very …