作者
Gautier Laurent, Nicolas Villani, Gabriela Hossu, Aymeric Rauch, Alain Noël, Alain Blum, Pedro Augusto Gondim Teixeira
发表日期
2019/8/1
期刊
European radiology
卷号
29
页码范围
4016-4025
出版商
Springer Berlin Heidelberg
简介
Objective
Evaluate and compare the image quality and acceptance of a full MBIR algorithm to that of an earlier full IR hybrid algorithm and filtered back projection (FBP).
Methods
Acquisitions were performed with a 320 detector-row CT scanner with seven different dose levels. Images were reconstructed with three algorithms: FBP, full hybrid iterative reconstruction (HIR), and a full model-based iterative reconstruction algorithm (full MBIR). The sensitometry, spatial resolution, image texture, and low-contrast detectability of these algorithms were compared. Subjective analysis of low-contrast detectability was performed. Ten radiologists answered a questionnaire on image quality and confidence in full MBIR images in clinical practice.
Results
The contrast-to-noise ratio of full MBIR was significantly higher than in the other …