作者
Pedro Augusto Gondim Teixeira, Anne-Sophie Formery, Gabriela Hossu, Daniel Winninger, Toufik Batch, Alban Gervaise, Alain Blum
发表日期
2017/2
期刊
European radiology
卷号
27
页码范围
437-446
出版商
Springer Berlin Heidelberg
简介
Propose
To establish evidence-based recommendations for musculoskeletal kinematic 4D-CT on wide area-detector CT.
Materials and methods
In order to assess factors influencing image quality in kinematic CT studies, a phantom consisting of a polymethylmethacrylate rotating disk with round wells of different sizes was imaged with various acquisition protocols. Cadaveric acquisitions were performed on the ankle joint during motion in two different axes and at different speeds to allow validation of phantom data. Images were acquired with a 320 detector-row CT scanner and were evaluated by two readers.
Results
Motion artefacts were significantly correlated with various parameters (movement axis, distance to centre, rotation speed and volume acquisition speed) (p < 0.0001). The relation between motion artefacts …