OBJECTIVE. The purpose of this study was to evaluate the image noise, low-contrast resolution, image quality, and spatial resolution of adaptive statistical iterative reconstruction in low-dose body CT.
MATERIALS AND METHODS. Adaptive statistical iterative reconstruction was used to scan the American College of Radiology phantom at the American College of Radiology reference value and at one-half that value (12.5 mGy). Test objects in low- and high-contrast and uniformity modules were evaluated. Low-dose CT with adaptive statistical iterative reconstruction was then tested on 12 patients (seven men, five women; average age, 67.5 years) who had previously undergone routine-dose CT. Two radiologists blinded to scanning technique evaluated images of the same patients obtained with routine-dose CT and low-dose CT with and without adaptive statistical iterative reconstruction. Image noise, low-contrast …