作者
Stefanie Mangold, Paola M Cannaó, U Joseph Schoepf, Julian L Wichmann, Christian Canstein, Stephen R Fuller, Giuseppe Muscogiuri, Akos Varga-Szemes, Konstantin Nikolaou, Carlo N De Cecco
发表日期
2016/6
期刊
European radiology
卷号
26
页码范围
1871-1878
出版商
Springer Berlin Heidelberg
简介
Purpose
To evaluate the impact of an advanced monoenergetic (ME) reconstruction algorithm on CT coronary stent imaging in a phantom model.
Materials and methods
Three stents with lumen diameters of 2.25, 3.0 and 3.5 mm were examined with a third-generation dual-source dual-energy CT (DECT). Tube potential was set at 90/Sn150 kV for DE and 70, 90 or 120 kV for single-energy (SE) acquisitions and advanced modelled iterative reconstruction was used. Overall, 23 reconstructions were evaluated for each stent including three SE acquisitions and ten advanced and standard ME images with virtual photon energies from 40 to 130 keV, respectively. In-stent luminal diameter was measured and compared to nominal lumen diameter to determine stent lumen visibility. Contrast-to-noise ratio was calculated.
Results …