作者
Vamsi Krishna Marothu, Rajendra N Dash, Saritha Vemula, Shravani Donkena, Ramesh Devi, Madhavi Gorrepati
发表日期
2012/12/1
期刊
Journal of pharmaceutical analysis
卷号
2
期号
6
页码范围
431-436
出版商
Elsevier
简介
An isocratic stability indicating RP-HPLC–UV method is presented for the determination of metaxalone (MET) in the presence of its degradation products. The method uses Dr. Maisch C18 column (250 mm×4.6 mm, 5 μm) with mobile phase consisting of acetonitrile–potassium dihydrogen orthophosphate buffer with 4 mL of 0.4% triethyl amine (pH 3.0; 10 mM) (58:42, v/v) at a flow rate of 1.0 mL/min. pH of the buffer was adjusted with o-phosphoric acid. UV detection was performed at 225 nm. The method was validated for specificity, linearity, precision, accuracy, limit of detection, limit of quantification and robustness. The calibration plot was linear over the concentration range of 1–100 μg/mL having a correlation coefficient (r2) of 0.999. Limits of detection and quantification were 0.3 and 1 μg/mL, respectively. Intra-day and inter-day precision (% RSD) was 0.65 and 0.79 respectively. The proposed method was used …
引用总数
2013201420152016201720182019202020212022202320242211123111