作者
Kai-Pun Wong, Brian Hung-Hin Lang, Sze-How Ng, Chung-Yeung Cheung, Christina Tin-Yan Chan, Chung-Yau Lo
发表日期
2013/12/1
期刊
Surgery
卷号
154
期号
6
页码范围
1158-1165
出版商
Mosby
简介
Introduction
Transcutaneous laryngeal ultrasonography (TLUSG) is a promising alternative to direct laryngoscopy in assessing perioperative vocal cord function. This study sought to evaluate the accuracy of TLUSG in assessing vocal cord function.
Methods
Altogether, 204 patients underwent TLUSG and direct laryngoscopy before and after elective thyroidectomy. For both examinations, vocal cord movements were independently graded. Grade I meant both vocal cords had normal movement; grade II meant ≥1 vocal cord had decreased movement; and grade III meant ≥1 vocal cord had no movement. Grade II or III on direct laryngoscopy was defined as vocal cord paresis or palsy (VCP). To assess accuracy, TLUSG findings were correlated with direct laryngoscopy findings.
Results
No patient had preoperative VCP, and 17 had unilateral postoperative VCP. The overall postoperative VCP rate was 5.1%. TLUSG …
引用总数
2013201420152016201720182019202020212022202320242118141714178151553