作者
Nguyen HB Ho, Dalton L Glasco, Jeffrey G Bell
发表日期
2022/8/2
期刊
ECS Sensors Plus
卷号
1
期号
2
页码范围
020601
出版商
IOP Publishing
简介
Benzalkonium (BA+) chloride is one of the most common preservatives used in prescription-based and over-the-counter eye drops. Knowing the concentration of BA+ in eye drops is important for both quality control (at the pharmaceutical preparation stage) and human health (ocular toxicity has been linked to BA+ use). This paper describes the design and fabrication of a benzalkonium-selective potentiometric sensor for the determination of BA+ in ophthalmic solutions. The sensor is composed of a 3D-printed ion-selective membrane (ISM) that selectively measures BA+ in the presence of potentially interfering ions routinely found in ophthalmic formulations (ie, Mg 2+, Ca 2+, Na+ and K+). The 3D printed BA+-ion-selective electrodes (ISEs) produced a Nernstian response of 55 mV/Decade across a range of 1.0 mM to 31.0 μM BA+ along with an LOD of 8 μM, which covers the relevant concentration range found in …