Yoshiki Yamaya, Harm J Bogaard, Peter D Wagner, Kyuichi Niizeki, Susan R Hopkins
发表日期
2002/1/1
期刊
Journal of Applied Physiology
卷号
92
期号
1
页码范围
162-168
出版商
American Physiological Society
简介
During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously compared Nellcor N-395 (Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10), a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt digit sensor (Ivy) to arterial blood oxygen saturation (SaO2 ) by cooximetry. Nine normal subjects, six athletes, and four patients with chronic disease exercised to maximum oxygen consumption (V˙o 2 max) under various conditions [normoxia, hypoxia inspired oxygen fraction (Fi O2 ) = 0.125; hyperoxia, Fi O2 = 1.0]. Regression analysis for normoxia and hypoxic data was performed (n = 161 observations, SaO2 = 73–99.9%), and bias (B) and precision (P) were calculated. RS10 offered greater validity than the other two devices …