David R Holmes III, Joel G Fletcher, Anja Apel, James E Huprich, Hassan Siddiki, David M Hough, Bernhard Schmidt, Thomas G Flohr, Richard Robb, Cynthia McCollough, Michael Wittmer, Christian Eusemann
发表日期
2008/12/1
期刊
European journal of radiology
卷号
68
期号
3
页码范围
409-413
出版商
Elsevier
简介
Dual-energy CT scanning has significant potential for disease identification and classification. However, it dramatically increases the amount of data collected and therefore impacts the clinical workflow. One way to simplify image review is to fuse CT datasets of different tube energies into a unique blended dataset with desirable properties. A non-linear blending method based on a modified sigmoid function was compared to a standard 0.3 linear blending method. The methods were evaluated in both a liver phantom and patient study. The liver phantom contained six syringes of known CT contrast which were placed in a bovine liver. After scanning at multiple tube currents (45, 55, 65, 75, 85, 95, 105, and 115mAs for the 140-kV tube), the datasets were blended using both methods. A contrast-to-noise (CNR) measure was calculated for each syringe. In addition, all eight scans were normalized using the effective …