作者
János Nemcsik, József Egresits, Taha El Hadj Othmane, Bertalan Csaba Fekete, Erzsébet Fodor, Tamás Szabó, Zoltán Járai, Csaba Jekkel, István Kiss, András Tislér
发表日期
2009/7/20
期刊
Kidney and Blood Pressure Research
卷号
32
期号
3
页码范围
223-229
出版商
S. Karger AG
简介
Background: Measuring arterial stiffness (augmentation index (AI), aortic pulse wave velocity (PWV)) in hemodialysis (HD) patients has prognostic significance. To assess its validity, the new oscillometric Arteriograph device (AIA, PWVA) was compared to the validated PulsePen tonometer (AIP, PWVP). Methods: AI and PWV were measured in 98 patients with both devices before HD. Validity was evaluated by Pearson’s correlation, Bland-Altman analysis, and by assessing the prognostic value of AI and PWV to predict cardiovascular (CV) mortality over 29 months. Results: Correlation between AIP and AIA was significant (R = 0.527, p < 0.001). The mean difference of AI values obtained by the two devices was –20.6%, and 30% of the paired AI differences fall outside the ±1 SD boundary of the mean between-device difference. There was no significant correlation between the PWVP and PWVA readings …