K Bredahl, A Long, M Taudorf, L Lönn, L Rouet, R Ardon, H Sillesen, JP Eiberg
发表日期
2013/5/1
期刊
European Journal of Vascular and Endovascular Surgery
卷号
45
期号
5
页码范围
450-455
出版商
WB Saunders
简介
OBJECTIVES
Volume estimation is more sensitive than diameter measurement for detection of aneurysm growth after endovascular aneurysm repair (EVAR), but this has only been confirmed on three-dimensional, reconstructed computer tomography (3-D CT). The potential of 3-D ultrasound (3-D US) for volume estimation in EVAR surveillance is unknown.
DESIGN
Prospective validation study comparing 3-D US with 3-D CT, using 3-D CT as the gold standard.
MATERIALS AND METHODS
From August 2011 to March 2012, 93 consecutive EVAR patients were enrolled and examined with both 3-D US and CT angiography (CTA). Image data were analysed in a mutual blinded setup using a 3-D interactive segmentation technique.
RESULTS
The technical success rate of 3D-US was 98% (91/93). In 91 EVAR patients (F/M; 10/81) eligible for further analysis, the mean maximum volume (SD) was 126 (58) ml using 3 …