作者
V Uppal, S Gangaiah, G Fletcher, J Kinsella
发表日期
2009/12/1
期刊
British journal of anaesthesia
卷号
103
期号
6
页码范围
882-885
出版商
Oxford University Press
简介
Background
The i-gel differs from other supraglottic airway devices, in that it has a softer, non-inflatable cuff. This study was designed to compare the performance of the i-gel and the LMA-Unique (LMA-U) when used during anaesthesia in paralysed patients.
Methods
Both devices were studied in 39 anaesthetized, paralysed patients in a randomized crossover trial. The primary outcome was airway leak pressure. Secondary outcomes included time to insertion, the number of insertion and reposition attempts, leak volumes, and leak fractions.
Results
There was no significant difference between the airway leak pressures of the two devices [median (IQR) leak pressures 25 (22–30) vs 22 (20–28) cm H2O for the i-gel and LMA-U, respectively; P=0.083, 95% CI of the mean difference −0.32 to 4.88 cm H2O]. The median (IQR) insertion time for the i-gel was …
引用总数