作者
V Uppal, G Fletcher, J Kinsella
发表日期
2009/2/1
期刊
British journal of anaesthesia
卷号
102
期号
2
页码范围
264-268
出版商
Oxford University Press
简介
Background
The i-gel (Intersurgical Ltd) is a novel device that differs from other supraglottic airway devices in that it has a softer and a non-inflatable cuff. Our study was designed to assess whether the i-gel is suitable to provide pressure-controlled ventilation (PCV) during anaesthesia by measuring the gas leaks and comparing these values with that of the tracheal tube.
Methods
Twenty-five patients, ASA I–II, were recruited to the study. Patients received a standard anaesthetic technique followed by an initial placement of the i-gel. The lungs were then ventilated at three different pressures (15, 20, 25 cm H2O) using PCV. The difference between the inspired and expired tidal volumes was used to calculate the leak volume. The leak fraction was defined as the leak volume divided by the inspired tidal volume. Following these observations, the i-gel was removed and replaced with …
引用总数
200820092010201120122013201420152016201720182019202020212022202320241261513151919871211109472
学术搜索中的文章