作者
Tiffany Jane Dwyer, Jennifer Ailsey Alison, Zoe Jane McKeough, Mark Russell Elkins, Peter Thomas Patrick Bye
发表日期
2009/10/1
期刊
Respiratory medicine
卷号
103
期号
10
页码范围
1511-1517
出版商
WB Saunders
简介
OBJECTIVE
Determine the SenseWear Pro3 Armband (SWA) accuracy for estimating energy expenditure (EE) and step count during treadmill walking in cystic fibrosis (CF) compared to healthy adults.
HYPOTHESIS
SWA estimation of EE would be less accurate for CF, than for healthy subjects, due to interference with the SWA skin sensors caused by the high salt concentration in the sweat of CF subjects.
METHODS
17 CF (mean age 26yr; FEV1 54% predicted) and 17 age-matched control subjects walked slightly faster than their comfortable pace on a treadmill for 20min, whilst simultaneously wearing the SWA and breathing through an open-circuit indirect calorimetry (IC) system. Subjects' steps were manually counted.
RESULTS
Combined EE data from all subjects showed no significant difference in EE measured by IC (6.0±3.4kcalmin−1) compared to the SWA estimate (6.3±2.5kcalmin−1), however the SWA …