作者
Masatsugu Niwayama, Ling Lin, Jun Shao, Nobuki Kudo, Katsuyuki Yamamoto
发表日期
2000/12/1
期刊
Review of Scientific Instruments
卷号
71
期号
12
页码范围
4571-4575
出版商
American Institute of Physics
简介
The inhomogeneity of tissue structure greatly affects the sensitivity of tissue oxygenation measurement by reflectance near-infrared spectroscopy. In this study, we investigated the influence of a fat layer on muscle oxygenation measurement by in vivo tests and Monte Carlo simulation, and we propose a method for correcting the influence. In the simulation, a three-dimensional model consisting of the epidermis, dermis, fat, and muscle layers was used. In in vivo tests, measurement sensitivity was examined by measuring oxygen consumption of the forearm muscle and the peak-to-peak variation of oxygenation in periodic exercise tests on the vastus lateralis using a newly developed multisensor type of tissue oximeter. Fat layer thickness was also measured by ultrasonography. The correction curve of measurement sensitivity against fat layer thickness was obtained from the results of simulation and in vivo tests. The values of …
引用总数
200120022003200420052006200720082009201020112012201320142015201620172018201920202021202220232024464471233463857946535452