作者
Lisa WM Leung, Banu Evranos, Alexander Grimster, Anthony Li, Mark Norman, Abhay Bajpai, Zia Zuberi, Manav Sohal, Mark M Gallagher
发表日期
2019/11
期刊
Journal of Interventional Cardiac Electrophysiology
卷号
56
页码范围
205-211
出版商
Springer US
简介
Background
The use of remanufactured single-use devices (SUDs), including cardiac electrophysiology catheters, has become established in the USA and other health care systems but without much published scientific evaluation on the relative safety or efficacy of these devices. In the United Kingdom (UK), the use of remanufactured SUDs has not been routine. We performed a structured evaluation of the safety and efficacy of a remanufactured circular mapping catheter (Stryker® remanufactured Lasso NAV 2515) during its introduction in our centre.
Methods
We prospectively evaluated the performance of a remanufactured circular mapping catheter in 100 consecutive patients undergoing an AF ablation. Operator feedback was obtained, assessing the device appearance, ease of use and function. As an indirect measurement of efficacy, acute …
引用总数
20192020202120222023202412834
学术搜索中的文章
LWM Leung, B Evranos, A Grimster, A Li, M Norman… - Journal of Interventional Cardiac Electrophysiology, 2019