作者
Jens Eckstein, Michael T Koller, Markus Zabel, Dietrich Kalusche, Beat A Schaer, Stefan Osswald, Christian Sticherling
发表日期
2008/5/27
期刊
Circulation
卷号
117
期号
21
页码范围
2727-2733
出版商
Lippincott Williams & Wilkins
简介
Background— Defibrillator lead malfunction is a potential long-term complication in patients with an implantable cardioverter-defibrillator (ICD). The aim of this study was to determine the incidence and causes of lead malfunction necessitating surgical revision and to evaluate 2 approaches to treat lead malfunction.
Methods and Results— We included 1317 consecutive patients with an ICD implanted at 3 European centers between 1993 and 2004. The types and causes of lead malfunction were recorded. If the integrity of the high-voltage part of the lead could be ascertained, an additional pace/sense lead was implanted. Otherwise, the patients received a new ICD lead. Of the 1317 patients, 38 experienced lead malfunction requiring surgical revision and 315 died during a median follow-up of 6.4 years. At 5 years, the cumulative incidence was 2.5% (95% confidence interval, 1.5 to 3.6). Lead malfunction …
引用总数
200820092010201120122013201420152016201720182019202020212022202320246141416191381914139844363
学术搜索中的文章